Measurement principle: | Solvent extraction, NDIR analysis (Infrared spectroscopy) |
Measurement range: | 0mg/l ~ 200mg/l dynamic ranging
|
Resolution | 0 to 99.9 mg/L; 0.1 mg/L 100 to 200 mg/L; 1 mg/L |
Repeatability | 0 to 9.9 mg/L; ± 0.2 mg/L ± 1 digit 10.0 to 99.9 mg/L; ± 2 mg/L ± 1 digit 100 to 200 mg/L; ± 4 mg/L ± 1 digit |
Measurement | Automatic after introduction of the oily water sample or manual operation, selectable |
Calibration | Automatic zero and span calibration after the oily water sample is introduced to the instrument. |
Extraction Solvent | S-316 solvent |
Sample/Solvent Volume | 20 mL/10 mL |
Extraction Method | Built-in extractor for oily water samples |
Display | Measured value; 3 digits LCD with backlight Message; Character display LCD with back-light (16×2 characters) |
Functions | Data memory up to 50 data, Self diagnostics, Stable type display, Interactive operation, Calendar clock |
Output | Digital display, RS-232C and parallel printer port |
Ambient Temperature | 0 to 40 ℃ |
Power | 100 to 240 V AC ±10%, 50/60 Hz 100 to 120 V AC; approx. 80 VA 200 to 240 V AC; approx. 120 VA |
Dimensions | 200(W)×315(D)×342(H) mm 7.8(W)×12.4(D)×13.5(H) in |
Mass | Approx. 7 kg/15.5 lb |